Model Number 13827
Device Problem No Flow (2991)
Patient Problem Inadequate Pain Relief (2388)
Event Date 03/18/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4). This malfunction is being reported because it is associated with an ongoing corrective action.
 
Event Description
It was reported that error codes were displayed on the latest software version of the clinician programmer upon initial inquiry of the pump. This software anomaly unexpectedly stopped the pump. The patient experienced inadequate pain relief. The error codes were able to be cleared, there are no further issues with the pump and the patient is doing fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROMETRA II PROGRAMMABLE PUMP
Type of DeviceIMPLANTABLE INFUSION PUMP
Manufacturer (Section D)
FLOWONIX MEDICAL, INC.
500 international drive
suite 200
mount olive NJ 07828
Manufacturer Contact
karen matis
500 international drive
suite 200
mount olive, NJ 07828
9734269229
MDR Report Key8458226
MDR Text Key140140463
Report Number3010079947-2019-00043
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Remedial Action Recall
Type of Report Initial
Report Date 03/18/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/20/2019
Device Model Number13827
Device Catalogue Number13827
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-