• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FLOWONIX MEDICAL, INC. PROMETRA II PROGRAMMABLE PUMP; IMPLANTABLE INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FLOWONIX MEDICAL, INC. PROMETRA II PROGRAMMABLE PUMP; IMPLANTABLE INFUSION PUMP Back to Search Results
Model Number 13827
Device Problem No Flow (2991)
Patient Problem Pain (1994)
Event Date 02/20/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This malfunction is being reported because it is associated with an ongoing corrective action.
 
Event Description
It was reported that error codes were displayed on the latest software version of the clinician programmer upon initial inquiry of the pump.This software anomaly unexpectedly stopped the pump.The patient experienced increased pain.The error codes were able to be cleared, there are no further issues with the pump and the patient is doing fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMETRA II PROGRAMMABLE PUMP
Type of Device
IMPLANTABLE INFUSION PUMP
Manufacturer (Section D)
FLOWONIX MEDICAL, INC.
500 international drive
suite 200
mount olive NJ 07828
Manufacturer Contact
karen matis
500 international drive
suite 200
mount olive, NJ 07828
9734269229
MDR Report Key8458228
MDR Text Key140151206
Report Number3010079947-2019-00039
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00810335020228
UDI-Public0100810335020228102451417190118
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/18/2019
Device Model Number13827
Device Catalogue Number13827
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/20/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/18/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-