The anesthesia workstation (system) was examined by our field service engineer.No fault was found.The nozzle units in the gas modules were replaced as a precaution.The system was returned back for clinical use, no further issues have been reported.The returned nozzle units were simulated use tested in a reference system.A successful system checkout (sco) was performed and thereafter a ventilation test, with parameter settings as during the reported event, was performed during 2 days without any deviation in the performance noticed.The reported deviation between set and delivered volume, was not reproduced and no alarms for high pressure were generated.Evaluation of the system device logs show that the treatment was performed in a volume controlled mode with tidal volume set to 430 ml, respiratory rate set to 14 b/min and expiratory minute volume upper alarm limit set to 15 l/min.A few alarms for expiratory minute volume low, respiratory rate high, airway pressure high and leakage were generated during the treatment.No alarms for expiratory minute volume high were generated.The log shows that the last sco before the event passed and there is no technical error in the log to indicate a system malfunction.Our conclusion, even though no fault was reproduced during simulated use test, is that one of the nozzle units caused an oscillation in one of the gas modules.Oscillations may result in an extra flow delivered from the gas module.Alarms will be activated if the set alarm limits are exceeded.
|