Model Number N/A |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/01/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source, foreign - event occurred in (b)(6).Recall zfa2018-00374 was performed on the affected item.This item was not delivered to the usa.The investigation is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during the workshop with surgeons in order to know how to use the instrument, in some items defects have been detected.The avantage straight impact plate got stuck in the curved handle the avantage straight impact plate thread as well as the curved shaft thread are defective and cannot be disassembled no patient or healthcare professionals have been involved.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Recall was performed on the affected item.Please note that the items in the scope of this recall were not delivered to the usa.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.Complaint sample was evaluated and investigation on the product did not show any non-conform results.According to the available data, the root cause of the event is related to design.Corrective action was initiated to investigate the described event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It has been reported that during the workshop with surgeons in order to know how to use the instrument, the avantage straight impact plate got stuck in the curved handle.The avantage straight impact plate thread as well as the curved shaft thread are malfunctioned and cannot be disassembled.No patient or healthcare professionals have been involved.
|
|