Model Number N/A
Device Problem Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source, foreign - event occurred in (b)(6).Recall zfa2018-00374 was performed on the affected item.This item was not delivered to the usa.The investigation is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during the workshop with surgeons in order to know how to use the instrument, in some items defects have been detected.The avantage straight impact plate got stuck in the curved handle the avantage straight impact plate thread as well as the curved shaft thread are defective and cannot be disassembled no patient or healthcare professionals have been involved.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Recall was performed on the affected item.Please note that the items in the scope of this recall were not delivered to the usa.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.Complaint sample was evaluated and investigation on the product did not show any non-conform results.According to the available data, the root cause of the event is related to design.Corrective action was initiated to investigate the described event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It has been reported that during the workshop with surgeons in order to know how to use the instrument, the avantage straight impact plate got stuck in the curved handle.The avantage straight impact plate thread as well as the curved shaft thread are malfunctioned and cannot be disassembled.No patient or healthcare professionals have been involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANTAGE STRAIGHT IMPACT PLATE
Type of Device
AVANTAGE STRAIGHT IMPACT PLATE
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
MDR Report Key8464410
MDR Text Key143814065
Report Number3006946279-2019-00197
Device Sequence Number1
Product Code HRS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Remedial Action Modification/Adjustment
Type of Report Initial,Followup
Report Date 05/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number110028874
Device Lot Number191750
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2018
Was the Report Sent to FDA? No
Date Manufacturer Received05/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A