Device evaluated by mfr.: the device was returned for analysis.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed a kink on the hypotube 87cm from the hub.The inflation lumen is stretched 3cm from the tip and is approximately 8.5cm long.The guidewire lumen is stretched 3mm from the tip and is approximately 11cm long.Microscopic examination revealed a longitudinal tear in the balloon 4mm from the tip and approximately 1mm long.There is a pattern on the balloon leading to the hole in the balloon.The marker band appears to be more distal than normal.The balloon is loosely folded.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the hypotube is kinked.
|
Reportable based on device analysis completed on (b)(6) 2019 it was reported that shaft kink occurred.The 88% stenosed, 21mmx2.0mm target lesion was located in the moderately tortuous and calcified left anterior descending artery.A 2.00mm x 20mm maverick balloon catheter was selected for use.However, it was noted that the balloon delivery shaft got kinked 80cm from the physician's hand to the kink point.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed balloon torn longitudinal.
|