Medtronic received a report of pipeline flex pushwire break.
The patient was undergoing pipeline flex placement in the treatment of a small unruptured saccular right internal carotid artery aneurysm, measuring 7mmx3mm, landing zone distal 3.
9mm, proximal 4.
0mm.
The vessel was normal tortuous.
The devices were prepared as indicated in the ifu.
A continuous flush was maintained during the procedure.
It was reported that during the procedure, the pipeline flex became stuck in the microcatheter.
The pipeline could not be pushed frmo the microcatheter, leading to stent and ewire cracked.
The stent was released in microcatheter and then withdrawn from the patient.
The broken segment of the pushwire removed from the patient.
There was reportedly no damage to the catheter.
There were no patient symptoms or complications associated with this event.
|