Model Number 1006-9321-000
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The reported event was addressed by the distributor performing a checkout and adjustment of the gas proportioning system.No further resolution information available.
 
Event Description
This report summarizes 1 malfunction event.A review of the event indicated that model 1006-9321-000 anesthesia gas machine experienced the gas proportioning system not operating as expected, causing the potential of a hypoxic mix of gas delivery.This report was from a single source.This event did not involve a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESTIVA 7100
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key8477850
MDR Text Key145279875
Report Number2112667-2019-00044
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial
Report Date 04/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/03/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator No Information
Device Model Number1006-9321-000
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/01/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-