Catalog Number 1014252-040 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 03/05/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that during device preparation, the 2.0x40mm armada 18 balloon leaked saline.There was no patient involvement and no reported clinically significant delay in the procedure.Another armada 18 was used to successfully complete the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
Internal file number - (b)(4).Evaluation summary: visual and functional inspections ware performed on the returned device.The reported inflation issue and leak were confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The investigation determined that the leak appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
|
|