Medtronic received a report that there was resistance when pushing the pipeline inside the middle and distal parts of the catheter due to moderate vessel tortuosity.
As a result the pipeline's pushwire was broken after 2 proximal markers, just after deployment of the pipeline.
However, it was confirmed that the pipeline was deployed without issue.
The patient was undergoing surgery for treatment of a posterior communicative aneurysm with a saccular morphology and a max diameter of 4.
5mm, neck diameter 2mm.
Dual antiplatelet treatment was administered with a pru level "as usual.
" post angiographic imaging showed that the aneurysm was successfully treated.
It was confirmed that there was no damage to the catheter seen.
There were no related patient symptoms.
|