STRYKER TRAUMA KIEL TIBIAL NAIL, STANDARD T2 TIBIA Ø9X345 MM ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
Catalog Number 18220934S |
Device Problems
Activation, Positioning or SeparationProblem (2906); Material Deformation (2976)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/22/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was not returned.
If additional information becomes available, it will be provided in a supplemental report.
Device disposition unknown.
|
|
Event Description
|
It was reported that, during a tibial nail procedure, when it came time to attach the nail to the holding bolt, the threads inside the nail would not engage with the holding bolt.
Multiple attempts with no success.
A second nail was opened and the nail successfully engaged with the holding bolt.
Two minute surgical delay reported.
No adverse consequences were reported.
|
|
Search Alerts/Recalls
|
|
|