Model Number N/A
Device Problem Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per company standard operating procedure since the device manufacture date is greater than one year from the event date.A getinge service territory manager (stm) was dispatched to evaluate the iabp.The stm was unable to reproduce the reported issue.A full calibration was performed and all functional and safety tests were performed and passed to meet factory specifications, and the iabp was returned to the customer and cleared for clinical service.
 
Event Description
While the getinge service territory manager (stm) was on-site, the facility biomed notified the stm that the cs300 intra-aortic balloon pump (iabp) was found in the facility hallway and reported that the iabp malfunctioned; it is unclear which type of malfunction occurred.It is unknown under which circumstances this event occurred; and it is unknown whether there was patient involvement; however no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key8483793
MDR Text Key141035922
Report Number2249723-2019-00535
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/12/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-