While walking by patient's room, nurse heard a trickling noise.When she looked into the room, she saw that the tube feed had become disconnected at the insertion site and tube feed was pouring down the iv pumps and onto the floor.Upon further inspection it appears that the tube became dislodged from the insertion spike.Nursing staff also reports multiple occasions recently where the feeding pump tubing begins leaking through a puncture hole, generally near the purple connector.Packaging was not saved for the two sets of leaking tubing that were saved.However, most likely the tubing sets come from lot # 190040039.
|