BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
|
Back to Search Results |
|
Model Number 6532 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/19/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that shaft damage occurred.The target lesion was located in the left anterior descending artery.A 15mm x 2.75mm nc quantum apex balloon catheter was used for dilatation.However, during the start of inflation, the shaft ruptured and blood went back into the indeflator.The device was completely removed and the procedure was completed using a new wire and a 2.5 x 15 nc quantum aprex balloon catheter.No patient complications were reported.
|
|
Event Description
|
It was reported that shaft damage occurred.The target lesion was located in the left anterior descending artery.A 15mm x 2.75mm nc quantum apex balloon catheter was used for dilatation.However, during the start of inflation, the shaft ruptured and blood went back into the indeflator.The device was completely removed and the procedure was completed using a new wire and a 2.5 x 15 nc quantum aprex balloon catheter.No patient complications were reported.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: returned product consisted of a nc quantum apex balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the hypotube, inflation lumen, and guidewire lumen.The hypotube is separated 73.9cm from the hub and appears to have been kinked prior to separation.Microscopic examination revealed no additional damages.There is blood present in the inflation lumen, guidewire lumen, and hub.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis found a separation of the hypotube.
|
|
Search Alerts/Recalls
|
|
|