BOSTON SCIENTIFIC CORPORATION NC EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7311
Device Problem Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/13/2019
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a nc emerge balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the hypotube, inflation lumen, core wire, and guidewire lumen.There is a longitudinal tear in the inflation lumen 19mm from the tip and is approximately 12mm long.Microscopic examination revealed no additional damages.There is blood present in the hub, inflation lumen, and balloon.The balloon is loosely folded.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event Description
Reportable based on device analysis completed on 21mar2019.It was reported that tip damage occurred.During preparation of a 4.00mm x 8mm nc emerge balloon catheter, it was noticed that the tip was damaged.The procedure was completed with another of the same device.No patient complications were reported.However, returned device analysis revealed shaft tear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC EMERGE
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8501134
MDR Text Key141578606
Report Number2134265-2019-03682
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeEI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/30/2020
Device Model Number7311
Device Catalogue Number7311
Device Lot Number22868850
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-