Model Number W4TR06
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/26/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during the implant procedure, the physician had difficulty engaging the setscrew to tighten to the lead pin. The physician removed the torque wrench and noted that the setscrew was still mounted on the torque wrench. The device was attempted and not used. A different device was used. No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed. Returned product analysis was performed and no anomalies were found. If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSOLARA QUAD CRTP
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8503668
MDR Text Key141591271
Report Number9614453-2019-01200
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 05/23/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/14/2020
Device Model NumberW4TR06
Device Catalogue NumberW4TR06
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/16/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-