• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number U128
Device Problems Signal Artifact/Noise (1036); High impedance (1291); Failure to Sense (1559)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/11/2019
Event Type  malfunction  
Event Description
It was reported that during the implant procedure this cardiac resynchronization therapy pacemaker (crt-p) system exhibited no sensing and high out of range pacing impedances on the left ventricular (lv) lead, measuring greater than 2000 ohms.The sensitivity was increased and the device is now sensing and pacing appropriately.This crt-p system remains in service.No adverse patient effects were reported.
 
Event Description
Additional information was received which indicated that the lv lead exhibited noise.Boston scientific technical services (ts) discussed troubleshooting and reprogramming options.The crt-p and lv lead remain in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE X4 CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key8504308
MDR Text Key141609248
Report Number2124215-2019-03524
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559402
UDI-Public00802526559402
Combination Product (y/n)N
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 07/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/24/2020
Device Model NumberU128
Device Catalogue NumberU128
Device Lot Number732498
Was Device Available for Evaluation? No
Date Manufacturer Received05/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age65 YR
-
-