• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INVIVE; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INVIVE; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number V173
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/09/2019
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) experienced a reset and went into safety core.The cause of the reset was unknown.Subsequently, a revision procedure occurred and the crt-p was explanted and replaced.No additional adverse patient effects were reported.
 
Event Description
This supplemental report is being filed as the device evaluation was completed.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode and that brady therapy remained available.Review of device memory identified an error.The error resulted in software resets performed in an attempt to correct an identified memory inconsistency.The corrupted memory was corrected in the laboratory and the device reverted to normal operation.The cause of the memory inconsistency was not able to be determined through laboratory testing.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode and that brady therapy remained available.Review of device memory identified an error.The error resulted in software resets performed in an attempt to correct an identified memory inconsistency.The corrupted memory was corrected in the laboratory and the device reverted to normal operation.The cause of the memory inconsistency was not able to be determined through laboratory testing.The battery was sent for analysis after the initial analysis was completed.The device case was removed to facilitate inspection of the internal components and further testing.The battery was removed from the device and replaced with an external power source.The current drain was measured and found to be normal.Detailed analysis of the battery identified an internal short, which was caused by a tear in the cathode insulating tube.The short resulted in the reported clinical observations.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) experienced a reset and went into safety core.The cause of the reset was unknown.Subsequently, a revision procedure occurred and the crt-p was explanted and replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVIVE
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key8505059
MDR Text Key141630621
Report Number2124215-2019-04020
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526536632
UDI-Public00802526536632
Combination Product (y/n)N
PMA/PMN Number
P030005/S079
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/27/2015
Device Model NumberV173
Device Catalogue NumberV173
Device Lot Number103828
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2019
Date Manufacturer Received01/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age90 YR
-
-