Model Number U125
Device Problems Failure to Capture (1081); Defective Device (2588)
Patient Problems Syncope (1610); Dizziness (2194)
Event Date 02/05/2019
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) patient experienced a syncopal episode.Loss of capture (loc) was observed on the right ventricular (rv) lead and left ventricular (lv) lead during telemetry.Isometrics and pocket manipulations were performed and no noise was reproduced, and the impedances were stable around 400 ohms.Data analysis showed there were issues with the right atrial (ra) lead auto threshold tests due to atrial rate to high for a test to start and fusion issues.There were no recorded errors for the rv auto threshold test.It was noted over the past few weeks the patient has experienced an increased number of presyncopal episodes.The rv ring was common to both the rv lead and lv lead pacing pathway, and it is suspected there is an intermittent issue involving the rv ring.The crt-p was reprogrammed and loc hasn't occurred since.This crt-p system remains in service.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CLONMEL LIMITED
cashel road
,
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
,
saint paul, MN 55112
6515826168
MDR Report Key8509788
MDR Text Key141779035
Report Number2124215-2019-04532
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559389
UDI-Public00802526559389
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/06/2019
Device Model NumberU125
Device Catalogue NumberU125
Device Lot Number704937
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/05/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/23/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age26 YR
-
-