AV-TEMECULA-CT TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012274-12
Device Problems Material Separation (1562); Material Too Soft/Flexible (4007)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/22/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4). During processing of this complaint, attempts were made to obtain complete event, patient and device information. The device is expected to be returned for investigation. It has not yet been received. A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was performed to treat a moderately tortuous and non-calcified lesion in the left anterior descending (lad) coronary artery. A 3. 0 x 12 mm trek rx balloon dilatation catheter (bdc) was used and the proximal shaft snapped and separated in two pieces as the bdc was being advanced without resistance over the unspecified guide wire and into the unspecified guide catheter. Both pieces of the bdc were retrieved from the patient anatomy through the guide catheter without resistance, and no part of the bdc remained in the patient anatomy. The physician noted that the shaft of the bdc felt flimsy during use. A non-abbott bdc was used to complete the procedure. There were no adverse patient effects, and there was no reported clinically significant delay in the procedure. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameTREK CORONARY DILATATION CATHETER
Type of DeviceCORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key8515432
MDR Text Key142139122
Report Number2024168-2019-02930
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/07/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date12/31/2021
Device Catalogue Number1012274-12
Device Lot Number90118G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/28/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-