• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS VOLISTA; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS VOLISTA; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Catalog Number ARD568811960
Device Problem Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/08/2019
Event Type  malfunction  
Manufacturer Narrative
The issue is being investigated by manufacturing site.(b)(4).
 
Event Description
On (b)(6) 2019 maquet sas became aware of an issue with one of the surgical lights- volista.As it was stated, there was a crack on the fork of the device which might lead to paint particles falling off.There was no injury reported, however it was decided to report the issue in abundance of caution as any paint particles falling off into sterile field or during procedure might be a source of contamination.Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
Getinge usa sales, llc (importer) is submitting this report on behalf of the legal manufacturer of the device (b)(6) exemption # e2018005.(b)(6).Contact person: (b)(6).Maquet sas became aware of a customer¿s problem with the volista device.As it was stated by the technician, there were cracks in the paint observed, specifically at the arc of the fork.There was no injury to a patient reported, however it was decided to report this issue based on the potential and in abundance of caution, as any particle falling from the device into the sterile field might be a source of contamination.When reviewing similar reportable events registered for volista surgical light we were able to find several similar issues compared to the problem investigated herein.In none of the complaints a serious injury or death occurred.Based on the information collected to date it was established that when the event occurred, the surgical light did not meet the manufacturer¿s specification.The most probable root cause that the cracks appear was the detachment of the bonded parts of fork.It was established that the cracks were located only on the outer coating and there was no impact to the internal structure of the fork assembly.The breaks on the coating correspond to the excessive gap between two mechanical parts under the outer material.We have no information regarding the exact time when the defect first appeared or if it was being used for patient treatment in the time when the event occurred.A corrective/preventive action investigation into the issue has revealed that the issue may be the result of several factors, summarized as issues with the previous understanding of post-gluing aspects.A design change improved the assembling methods from glue-bonding to welding of the parts in production, since beginning of 2017.In addition, the device was included in the scope of field action msa-2019-001-iu and it was decided that the defective part will be replaced, therefore getinge does not propose any other action at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLISTA
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
MDR Report Key8518569
MDR Text Key145753279
Report Number9710055-2019-00143
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Inspection
Type of Report Initial,Followup,Followup
Report Date 06/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberARD568811960
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-