It was reported that the patient¿s generator was at less than 25% after being implanted in 2017, and the physician believed that the patient¿s device may be prematurely depleting.Programming history was reviewed for the generator, and the settings on the day of implant were available.A battery life calculation was performed using the programming history and patient¿s current settings, and assumptions were made for any gaps if present.It was assumed that the patient's settings were changed halfway between the date of implant and the day the settings were given.There was 3.7 years predicted to near end of service.No anomalies were seen.Device history records were reviewed for the generator.The device passed all specifications prior to distribution, and the device was not laser routed, and would not be susceptible to premature depletion due to laser routing.The patient had not had any electrical current that may have damaged the generator pass through the body.Surgery is likely but has not occurred to date.No additional or relevant information has been received to date.
|