It was reported that the patient presented for follow-up.
Upon interrogation, the pulse generator exhibited a false elective replacement indicator alert on (b)(6) 2014, this was before the date of implant.
It was noted that the patient had not been seen since the implant procedure.
The device being exposed to electrocautery during the implant procedure, was suspected as the cause.
The device had triggered end of service on (b)(6) 2018.
After a firmware download, the device resumed normal function.
There were no consequences to the patient and the patient was stable post event.
|