• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Device Problems Failure to Fold (1255); Stretched (1601); Material Twisted/Bent (2981)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/08/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported the intraocular lens (iol) in a preloaded delivery system had a stretched haptic, was twisted, and did not fold as desired.There was no patient contact and a backup lens was implanted without issue.Additional information was requested.
 
Manufacturer Narrative
The device with the lens was returned.The plunger is oriented correctly.Viscoelastic is observed in the device.The plunger is at the trailing optic edge.The lens is advanced to mid-nozzle.The trailing haptic is folded in on the optic.The leading haptic is extended.Product history records were reviewed and the documentation indicated the product met release criteria.A qualified viscoelastic was observed.No problems are observed with the returned device.The plunger, lens and haptic positions are acceptable.Straight leading haptics are not a product malfunction.This is an acceptable position per the diagrams provided in the dfu.A straight leading haptic may occur: due to the normal folding variations as indicated in the dfu diagrams.If the initial plunger advancement is faster than the recommended target, the leading haptic may be forced past the internal folding feature.If the ovd fill rate is too fast the folding effect on the leading haptic is minimized.If there is excessive delay between the device preparation and subsequent delivery the leading haptic may start to unfold.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If the operating room temperature is too high (> 23°c / 73° f) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement or contribute to incorrect haptic folding.Any of the above listed causes alone, or in combination, may create the reported event.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Event Description
Additional clarification received reports the stretched haptic was the leading haptic.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
MDR Report Key8524764
MDR Text Key142472656
Report Number1119421-2019-00489
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 07/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2020
Device Model NumberAU00T0
Device Lot Number12577827
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2019
Date Manufacturer Received07/27/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PROVISC
-
-