STRYKER ORTHOPAEDICS-MAHWAH UNKNOWN ACCOLADE TMZF HIP STEM HIP IMPLANT Back to Search Results
Catalog Number UNK_SHC
Device Problems Degraded (1153); Device Dislodged or Dislocated (2923)
Patient Problems Inflammation (1932); Injury (2348); Reaction (2414)
Event Date 03/22/2019
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation. Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event Description
It was reported that the patient's left hip was revised due to disassociation of the head from the stem. Intraoperatively, trunnionosis, corrosion, "a lot of black tissue", and liner wear at the rim were noted. A 36 +10 lfit v40 head, accolade tmzf stem, and poly liner were revised to an eon stem, ceramic femoral head, and mdm/adm liner construct.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameUNKNOWN ACCOLADE TMZF HIP STEM
Type of DeviceHIP IMPLANT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
Manufacturer Contact
collin neitzel
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key8526903
MDR Text Key142340606
Report Number0002249697-2019-01661
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Catalogue NumberUNK_SHC
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/16/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 04/18/2019 Patient Sequence Number: 1
-
-