Manufacturer evaluation results: visual inspection results: when the unit was unboxed a non-sechrist water trap was found in the packaging and it appeared that the unit has damage to the fio2 knob.
When the unit was tested the fio2 reading at.
21 was out of specification and the balance was also out of specification.
The fio2 reading at.
21 and the balance readings at.
30 and.
90 being above baseline are due to the fio2 knob being bent and loose, the unit has been dropped or hit against some unknown object and it does not appear that the unit was damaged from shipping.
User's manual states: the sechrist air/oxygen mixer is a "precision pressure regulation and proportioning device, which is designed to accurately mix medical grade air and medical grade oxygen (02).
Prior to each clinical usage, the user should perform an alarm test and analyze the full fio2 range.
With an accurately calibrated oxygen analyzer, the user should analyze the fio2 at the following settings; 21%, 40%, 60%, 80%, and 100%.
Additionally, the user should briefly disconnect one supply gas to assure that the bypass/alarm system is functioning.
With a single supply gas disconnected, the audible alarm should sound and the analyzed fio2 should indicate the fio2 of the single supply gas; i.
E.
21% if the oxygen was disconnected and 100% if the air supply was disconnected.
The user's manual also has a "warning statement "the user of the sechrist air/oxygen mixer shall have sole responsibility for any malfunction, which results from improper usage, faulty maintenance, improper and/or unauthorized repairs, damage or alteration performed by anyone other than sechrist industries.
" device history record review: mixer model 3500cp-g, s.
N.
(b)(4) was manufactured on 08/27/2018.
There is no indication that there were any relevant discrepancies during manufacturing.
A review of the device history record (dhr) found no non-conformance that could cause or contribute to the reported issue.
|
During a procedure, it was noticed that the mixer was not delivering the correct % of o2.
Mixer is part of a heart/lung machine.
Staff noticed immediately and switched to another mixer.
Patient was not injured.
Customer also reported that this same issue occurred on three other patients during different days of treatment and after the fourth occurrence, the customer decided to remove it from use and return it to sechrist for evaluation and or repair.
This is the second of four reports being filed for this device.
See mfr #'s 2020676-2019-00006, 2020676-2019-00008 & 2020676-2019-00009.
|