BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number U128
Device Problems Over-Sensing (1438); Pacing Problem (1439); Low impedance (2285); Device Sensing Problem (2917)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2018
Event Type  Injury  
Manufacturer Narrative
As no additional information regarding this event is expected, our investigation is complete.This record will be updated if additional information becomes available.
 
Event Description
It was reported that the right ventricular (rv) lead connected to this cardiac resynchronization therapy pacemaker (crt-p) was surgically abandoned.It was noted that pacing impedance on the rv lead had dipped below 200 ohms several times over the past year, resulting in an automatic lead safety switch to unipolar.Noise and oversensing were noted, resulting in failure to pace.The pacing inhibition caused pauses lasting 1 to 4 seconds.The rv lead was surgically abandoned and replaced.This crt-p remains in service with the patient's new rv lead.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALITUDE X4 CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CLONMEL LIMITED
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key8534180
MDR Text Key142597055
Report Number2124215-2019-07169
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559402
UDI-Public00802526559402
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/06/2019
Device Model NumberU128
Device Catalogue NumberU128
Device Lot Number716867
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/27/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age87 YR
-
-