Model Number 989605406772
Device Problem Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
 
Event Description
A customer reported encountering an articulation issue with their s7-3t model transducer during a tee (transesophageal) examination.There was no injury associated with this event.
 
Manufacturer Narrative
Evaluation of the transducer confirmed the articulation issue as described by the customer.Investigation of the device revealed cracks on the beading and deep scratches on the tip shell.Performance testing found the transducer would not rotate forward and failed color imaging tests.The damage found on the probe is indicative of improper transducer care and maintenance.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S7-3T - 989605406771
Type of Device
TRANSDUCER, ULTRASONIC, DIAGNOSTIC
Manufacturer (Section D)
PHILIPS ULTRASOUND, INC
22100 bothell everett highway
bothell WA 98021
MDR Report Key8536903
MDR Text Key142887975
Report Number3019216-2019-00017
Device Sequence Number1
Product Code ITX
UDI-Device Identifier00884838061439
UDI-Public(01)00884838061439
Combination Product (y/n)N
PMA/PMN Number
K043535
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 03/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number989605406772
Device Lot NumberB2HX90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2019
Was the Report Sent to FDA? No
Date Manufacturer Received03/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-