Catalog Number IAB-06840-U |
Device Problems
Deformation Due to Compressive Stress (2889); Difficult to Advance (2920)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/08/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that on insertion of the intra-aortic balloon (iab) the dilator with the sheath was inserted into the patient's femoral artery, the md was not able to insert.
The blood vessel is sufficiently dilated.
Therefore the dilator was removed, and the doctor found the end of the dilator was damage.
As a result, a new set was used.
The patient outcome reported as fine.
There was no report of patient complication or serious injury and death.
|
|
Event Description
|
It was reported that on insertion of the intra-aortic balloon (iab) the dilator with the sheath was inserted into the patient's femoral artery, the md was not able to insert.
The blood vessel is sufficiently dilated.
Therefore the dilator was removed, and the doctor found the end of the dilator was damage.
As a result, a new set was used.
The patient outcome reported as fine.
There was no report of patient complication or serious injury and death.
|
|
Manufacturer Narrative
|
(b)(4).
Teleflex received the device for investigation.
The reported complaint that the dilator tip was damaged is confirmed by visual inspection of the returned samples.
The appearance of the dilator tip damage is an inconsistent diameter with indents or pinches around the tip opening.
The root cause of the damaged dilator tip is undetermined.
A device history record (dhr) review was conducted for the lot number with no relevant findings.
The device passed all manufacturing specifications prior to release.
Teleflex assessed the risk for the reported complaint.
There are no new or revised risks.
This will be monitored for any developing trends.
No further action required at this time.
|
|
Search Alerts/Recalls
|