It was reported that the arctic sun device was insufficiently cooling.
The device was placed into manual control and the system diagnostics displayed the system hours 4786, pump hours 4316, t1 was 6.
2 c, t2 was 6.
3 c, and t4 was 4.
3 c.
Mss advised that the device was making cold water and that the pads may be improperly sized for the patient.
Per follow up, the patient was switched to an alternate device and was able to reach target and complete therapy.
The device was sent to biomed where it passed a functional check and was placed back into service.
|