• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - CARTAGO CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - CARTAGO CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8537
Device Problem Reflux within Device (1522)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an unspecified quantity of clearlink continu-flo solution sets had back check valves which "seem to be allowing the primary and secondary bags to mix".This was noted during product use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed to the photograph using the naked eye which revealed backflow on the device.The reported condition was verified.By the nature of the sample, no additional tests were performed.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - CARTAGO
cartago
MDR Report Key8547255
MDR Text Key143126719
Report Number1416980-2019-02223
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412048970
UDI-Public(01)00085412048970
Combination Product (y/n)N
PMA/PMN Number
K961225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2C8537
Device Lot NumberR18E30121
Was Device Available for Evaluation? No
Date Manufacturer Received05/22/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PRIMARY AND SECONDARY BAGS
-
-