• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. AUTOCAT2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. AUTOCAT2 WAVE; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAP-0500
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/01/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that after insertion of the intra-aortic balloon (iab), the staff were unable to get the pump to initiate pumping, each time they attempt to pump, they have the purge failure alarms.As a result, the iab was removed because there was no other pump available to support the patient.There was no report of patient complications, serious injury or death.
 
Manufacturer Narrative
(b)(4).Teleflex received the device for investigation.The reported complaint of "purge failure alarm" is confirmed.The inlet fitting and 1 psi release valve were noted damaged, which could result in the reported alarm.The root cause of how the pcs assembly got damaged is undetermined.A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.This will be monitored for any developing trends.No further action required at this time.
 
Event Description
It was reported that after insertion of the intra-aortic balloon (iab), the staff were unable to get the pump to initiate pumping, each time they attempt to pump, they have the purge failure alarms.As a result, the iab was removed because there was no other pump available to support the patient.There was no report of patient complications, serious injury or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT2 WAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key8549206
MDR Text Key143053976
Report Number3010532612-2019-00111
Device Sequence Number1
Product Code DSP
UDI-Device Identifier00801902051714
UDI-Public00801902051714
Combination Product (y/n)N
PMA/PMN Number
K060309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIAP-0500
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2019
Date Manufacturer Received05/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-