STERIS MEXICO, S. DE R.L. DE C.V. 16" EAGLE STERILIZER Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Burn, Thermal (2530)
Event Date 03/26/2019
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the sterilizer and found that the s3 valve had failed.The s3 valve prevents steam and water from entering the chamber of the unit.As the valve was not functioning properly, this allowed water and steam to enter the chamber of the unit subsequently causing the reported event to occur.The technician rebuilt the s3 valve, tested the unit, confirmed it to be operating according to specification, and returned it to service.A 3-year complaint review indicates this to be an isolated event.No additional issues have been reported.
 
Event Description
The user facility reported that an employee obtained a burn from hot water while operating their 16" eagle sterilizer.Medical treatment was sought and administered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
16" EAGLE STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, nuevo leon 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key8552025
MDR Text Key145140031
Report Number3005899764-2019-00043
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/26/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-