Model Number 9438-05
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/29/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a transmitter failed error occurred. No product or data was provided for evaluation. Confirmation of the allegation and a root cause could not be determined. No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4). Date received by manufacturer - correction - please note that the first report submitted under 3004753838-2019-36993 was submitted with an incorrect date. This follow-up report is to note that it should have reflected a date of (01-apr-2019).  .
 
Event Description
The product was evaluated. The device was visually inspected and passed. Voltage testing was performed and failed due to no voltage. No log data available for review. Confirmation of the allegation and root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of DeviceCONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key8552586
MDR Text Key144285395
Report Number3004753838-2019-36993
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation
Type of Report Initial,Followup
Report Date 04/25/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9438-05
Device Catalogue NumberSTT-GL-004
Device Lot Number5241693
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2019
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/13/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-