• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS VOLISTA LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS VOLISTA LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Catalog Number ARD568811960
Device Problem Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/16/2019
Event Type  Malfunction  
Manufacturer Narrative

The issue is being investigated by manufacturing site. (b)(4). ,.

 
Event Description

On (b)(4) 2019 maquet (b)(4) became aware of an issue with one of the surgical lights- volista. As it was stated, there was a crack on the fork of the device which might lead to paint particles falling off. There was no injury reported, however it was decided to report the issue in abundance of caution as any paint particles falling off into sterile field or during procedure might be a source of contamination. (b)(4).

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVOLISTA
Type of DeviceLIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
PASCAL JAY - MAQUET SAS
parc de limère
avenue de la pomme de pi
orléans cedex 2 45074
FR 45074
Manufacturer Contact
parc de limère
avenue de la pomme de pi
orléans cedex 2 45074
0332382587
MDR Report Key8555103
MDR Text Key143401625
Report Number9710055-2019-00155
Device Sequence Number1
Product Code FSY
Combination Product (Y/N)N
Reporter Country CodeSP
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 05/20/2019
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received04/26/2019
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator NO INFORMATION
Device Catalogue NumberARD568811960
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/16/2019
Was Device Evaluated By Manufacturer? No
Date Device Manufactured07/09/2014
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-