It was reported that the arctic sun device was displaying an extended period of cold water temp alarm.
The water temperature was between 4-9c, the target temperature was 37c, and the patient temperature was 37c.
The flow rate was marginal to good at 2lpm - 2.
8lpm.
The patient was on dialysis, and was administered tylenol, mag, buspar and antibiotics.
Per troubleshooting, it was explained that the patient's heat generation from shivering could contribute to the issue, and that the facility needed to be diligent with skin assessments.
Additional information was received from the facility on the 11apr2019, that the patient was able to complete therapy with the same set of pads.
|