• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS MERIT CUSTOM KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS MERIT CUSTOM KIT Back to Search Results
Catalog Number K04-MK2100
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/22/2019
Event Type  malfunction  
Manufacturer Narrative
The suspect device has been returned for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event Description
The distributor reported a defect in the packaging.This was identified during their initial inspection of received product.The device was not sent to a user facility.
 
Manufacturer Narrative
The suspect device was returned for evaluation.The complaint is confirmed.The root cause is attributed to the manufacturing process.A review of the device history record was performed and no exception documents were found.A review of the complaint database was performed and no similar complaints for this lot number were found.Corrective actions are in process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERIT CUSTOM KIT
Type of Device
CUSTOM KIT
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS
8830 siempre viva rd #100
san diego CA 92154
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS
8830 siempre viva rd #100
san diego CA 92154
Manufacturer Contact
katie swenson cqe,cba,cqpa.
1600 merit parkway
south jordan, UT 84095
8012531600
MDR Report Key8562690
MDR Text Key143680307
Report Number3011642792-2019-00007
Device Sequence Number1
Product Code OJA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date01/31/2022
Device Catalogue NumberK04-MK2100
Device Lot NumberI1508802
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received06/18/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-