• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP* APPLIER W/MECH STOP; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP* APPLIER W/MECH STOP; APPLIER, SURGICAL, CLIP Back to Search Results
Catalog Number LX107
Device Problems Peeled/Delaminated (1454); Failure to Form Staple (2579); Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date of event: only event year known: 2019.Device analysis: the analysis results confirmed that the lx107 device was returned nonfunctional as the jaws were misaligned; therefore, the clips could not be loaded into the jaws.It is possible that the damaged was due to an improper handling of the device.Please note that as stated in the ifu: "all surgical instruments are subject to a degree of wear and tear because of normal use.A regular and precise visual check of the instrument should be made before each use.The manufacturing record evaluation was performed and identified a issue related to the reported incident.The necessary actions to ensure the final product quality have been taken and documented in the appropriate quality system.The final quality release criteria were met before this batch was released for distribution.The certificate records are accessible through external manufacturing.
 
Event Description
It was reported that devices have plastic coating chipping off, some have the tips bent and some do not hold clips.There is no additional information.No patient consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8568983
MDR Text Key143884314
Report Number3005075853-2019-18639
Device Sequence Number1
Product Code GDO
UDI-Device Identifier10705036012856
UDI-Public10705036012856
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 03/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberLX107
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2019
Date Manufacturer Received04/08/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2014
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-