Cine analysis: the cine images were received.The images are of the targeted vessel anatomy in the right proximal popliteal artery.Images 1 and 3 appear to be duplicates of each other, above the knee, and both contain what appears to be a support sheath at the top of the image.Image 2 is at the knee and shows a tight stenosis of the artery.The chocolate device cannot be seen in any of the images.Device analysis: the chocolate device was received for evaluation.No ancillary devices were received for analysis.The chocolate device was received with the balloon chamber in a post-inflation profile.The distal end of the balloon chamber was examined and the chocolate nitinol cage remained firmly attached.The proximal end of the balloon chamber was examined and the chocolate nitinol cage remained firmly attached to the catheter.A 0.018¿ guidewire was loaded through the distal tip of the catheter and navigated out the guidewire port with ease.A 10cc water filled syringe was attached to the inflation port of the y-manifold and a vacuum was pulled.The vacuum could be maintained.A 20cc water filled syringe with manometer was attached to the inflation port and the syringe was inflated to a pressure of 12-atm, but the balloon would not inflate.The pressure was left on for over 20-minutes.The inflation lumen is most likely obstructed with solidified contrast from the procedure.The distal end of the returned device was compared to the distal end of another chocolate device that was not involved with this procedure.The distal nitinol wires are standing proud of the distal balloon bond on the returned device but the nitinol wires remain bonded to the chocolate device.If information is provided in the future, a supplemental report will be issued.
|