Catalog Number 115269
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/09/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during continuous renal replacement therapy (crrt) of a pediatric patient using a prismaflex control unit, communication error alarm code 2 was generated causing the prismaflex control unit to enter a patient safe state.Treatment was discontinued.The extracorporeal blood in the circuit clotted and was unable to be returned.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was not received for evaluation; therefore, a device analysis could not be completed.A device history review revealed no issues that could have caused or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - LUND MONITORS
lund skane lan
MDR Report Key8574978
MDR Text Key145655887
Report Number9616026-2019-00025
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
K131516
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/02/2019,06/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/02/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number115269
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA05/02/2019
Distributor Facility Aware Date04/09/2019
Event Location Hospital
Date Report to Manufacturer05/02/2019
Date Manufacturer Received05/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN PRISMAFLEX FILTER SET
-
-