DEPUY ORTHOPAEDICS, INC. 1818910 SIG FEM ADPT TORQUE WRENCH; KNEE INSTRUMENT : TORQUE DEVICES
|
Back to Search Results |
|
Catalog Number 961673 |
Device Problems
Device-Device Incompatibility (2919); Material Twisted/Bent (2981); Mechanical Jam (2983)
|
Patient Problem
Not Applicable (3189)
|
Event Date 04/08/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Dr.(b)(6) was using the torque wrench to tighten the 5mm adapter while doing a sigma knee revision.The adapter is stuck in the torque wrench despite multiple efforts to free the implant.It remains stuck in the torque wrench.We opened a second adapter.Was surgery delayed due to the reported event? --> yes, if yes, number of minutes: --> 5-10.Was procedure successfully completed? --> yes.Were fragments generated? --> no.Patient status/ outcome / consequences --> no.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : examination of the returned device confirmed the reported event.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: product complaint (b)(4).Investigation summary:examination of the returned device confirmed the reported event.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
|