BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND SYSTSEM CONSOLE (REMANUFACTURED); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
|
Back to Search Results |
|
Model Number 9770600Y |
Device Problems
Loss of Power (1475); Device Displays Incorrect Message (2591); Battery Problem (2885)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The manufacturer has received the sample and will be evaluated.Results are expected soon.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number dyasac025 showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per biomed: the unit shows inaccurate battery life, when the display shows twenty percent of battery life the scanner shuts off.
|
|
Manufacturer Narrative
|
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of system shutdown abruptly was confirmed.The scanner shuts down prematurely.The root cause of the failure was identified as an internal failure in the lithium ion battery.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested, and returned to the customer.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
|
|
Event Description
|
Per biomed: the unit shows inaccurate battery life, when the display shows twenty percent of battery life the scanner shuts off.
|
|
Event Description
|
Per biomed: the unit shows inaccurate battery life, when the display shows twenty percent of battery life the scanner shuts off.
|
|
Manufacturer Narrative
|
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of system shutdown abruptly was confirmed.The scanner shuts down prematurely.The root cause of the failure was identified as an internal failure in the lithium ion battery.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested, and returned to the customer.A lot history review (lhr) review is not possible, as the device is manufactured using a unique serial number and not by lot number.A history review of serial number (b)(4)showed no other similar product complaint(s) from this serial number.
|
|