Device evaluated by mfr.: received product consisted of a quantum maverick balloon catheter.The balloon was tightly folded with blood in the wire lumen.The blood in the wire lumen (inner shaft) was removed by soaking in a warm water bath with ultrasonic cleaning for 1 hour.The hypotube, inner/outer shafts, balloon and tip were microscopically and visually inspected.Inspection found no damage or irregularities with the returned device.The guidewire used in the procedure was not returned for analysis, so a.014" test guide wire was used for functional testing.The test wire was loaded into the tip of the quantum maverick and advanced out through the port/exit notch.There were no issues, and the wire performed in the expected fashion.
|