• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 150 - DEU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 150 - DEU Back to Search Results
Model Number 27053
Device Problems Premature Discharge of Battery (1057); Failure to Charge (1085); Failure to Power Up (1476)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to resmed and an evaluation confirmed the complaint.The internal battery was replaced to address this issue.The device was serviced and fully tested before it was returned to the customer.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.(b)(4).
 
Event Description
It was reported to resmed that an astral device failed to charge its internal battery and had an internal battery with a reduced level of capacity.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
The astral device was returned to resmed for an investigation.Review of the device data logs confirmed the reported complaint and revealed an external battery communications lost alarm.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the device failure to start up and external battery communications lost alarm were due to a software issue; the sf180 was due to device operation in a temperature outside of the device specification and the failure to charge its internal battery was due to operator error.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference #: (b)(4).
 
Event Description
It was reported to resmed that an astral device failed to start up, charge its internal battery and had an internal battery with a reduced level of capacity.It was reported during evaluation of the device by resmed, the astral device displayed an error message (sf180) related to a battery charger fault.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - DEU
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive.
bella vista
sydney, nsw 2153
AU  2153
MDR Report Key8584300
MDR Text Key144263470
Report Number3004604967-2019-00155
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
PMA/PMN Number
K152068
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 06/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27053
Device Catalogue Number27053
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/15/2019
Distributor Facility Aware Date05/19/2019
Device Age19 MO
Date Manufacturer Received05/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-