The investigation is still in progress.
Once the investigation is complete, a supplemental report will be filed.
The information provided by bd represents all of the known information at this time.
Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
The device was evaluated by a biomedical technician at the complainant's facility.
|
It was reported that the arcticsun device data displayed multiple alert 113s ( reduced water temperature control).
The water temperature stayed in the 20cs even though the patient data was above the target temp.
The device was placed in manual mode and the water temperature was set to 4c: t1 and t2 were 21.
8c, t4 was 3.
1c, the mixing pump command was running at 100%, and the circulation pump command was running at 47%.
|