Model Number 400CE
Device Problem Insufficient Heating (1287)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/08/2019
Event Type  malfunction  
Manufacturer Narrative
Gentherm received a complaint ((b)(4)) stating the device could not reach the desired temperature.The device was switched out and the procedure was completed successfully.No patient harm or injury occurred.Gentherm evaluated the device and found the high amp heater was defective.(b)(4).
 
Event Description
Customer reported the device did not reach desired temperature.
 
Manufacturer Narrative
Device returned for evaluation.Service evaluation confirmed the heater was defective.The heater was replaced to correct.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOTHERM 400CE
Type of Device
CARDIOPULMONARY BYPASS TEMPERATURE CONTROLLER
Manufacturer (Section D)
GENTHERM MEDICAL, LLC
12011 mosteller
cincinnati OH 45241
Manufacturer Contact
christina miracle
12011 mosteller
cincinnati, OH 45241
8009897373
MDR Report Key8593288
MDR Text Key150887551
Report Number1516825-2019-00004
Device Sequence Number1
Product Code DWC
UDI-Device Identifier10613031860221
UDI-Public(01)10613031860221
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122813
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number400CE
Device Catalogue Number86022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/08/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-