It was reported that patient had an appointment, the device was interrogated and the normal mode output current was set to 0.00 ma while the magnet mode current was still set to 2.00 ma.The np stated that the patient's device was not turned off at the time of the event in february.The facility cannot determine whether the patient's device has been turned off since february, they have decided to start the patient back up at lower levels and titrate the patient up slowly.She stated that diagnostics were run and were within normal limits.Programming history was reviewed and had settings from february and april appointments.It is seen that the last event in february was a programming event to 1.75ma and when the patient came in april the settings were at 0ma.There is no indication form the settings of a faulted diagnostic test that may have unintentionally changed settings.There is a two month gap where the patient may have potentially been seen by a different physician and programmed to these settings however this needs to be confirmed.Other than the note anomaly above, no other anomalies were noted.No additional relevant information has been received to date.
|