• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Material Twisted/Bent (2981)
Patient Problems Vascular System (Circulation), Impaired (2572); No Known Impact Or Consequence To Patient (2692)
Event Date 05/07/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryoablation procedure, the sheath was having difficulty passing through the puncture site when it was noticed the tip of the sheath was kinked.The sheath was replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the sheath 4fc12 with lot number 85063 was returned and analyzed.Visual inspection of the sheath showed a deformation at the tip of the shaft.Additionally, the shaft appeared to be kinked 2.4 inches from the tip.The functionality test showed the deflection mechanism of the sheath did not work per specification.In conclusion, the reported sheath tip kink was confirmed through testing.The sheath failed the return product inspection due to the shaft kink, and tip kink.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key8599610
MDR Text Key144710273
Report Number3002648230-2019-00346
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/22/2020
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number85063
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2019
Date Manufacturer Received06/02/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age68 YR
-
-