ABBOTT GERMANY ARCHITECT CA 19-9XR Back to Search Results
Catalog Number 02K91-32
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
All available patient information was included. Additional patient details are not available. An evaluation is in process. A final report will be submitted when the evaluation is complete.
 
Event Description
The customer observed falsely depressed ca19-9 results on the architect i2000sr analyzer. The following data was provided: initial 554 u/ml, repeated on alinity 920 u/ml. And had a previous result of 1020 u/ml. There was no impact to patient management reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCHITECT CA 19-9XR
Type of DeviceCA 19-9
Manufacturer (Section D)
ABBOTT GERMANY
max-planck-ring 2
wiesbaden 65205
GM 65205
Manufacturer (Section G)
ABBOTT GERMANY
max-planck-ring 2
wiesbaden 65205
GM 65205
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key8605166
MDR Text Key144879779
Report Number3002809144-2019-00285
Device Sequence Number1
Product Code NIG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K052000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 05/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date09/13/2019
Device Catalogue Number02K91-32
Device Lot Number93018M800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/29/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

Patient Treatment Data
Date Received: 05/13/2019 Patient Sequence Number: 1
Treatment
ARCHITECT I2000SR ANALYZER; LN: 03M74-01, SN: (B)(4)
-
-