During use, an optease sheath was attempted to be inserted, however the marker was scratch (frayed).
There was no patient injury.
Therefore the optease was replaced with a new device and the procedure was completed.
The target lesion was not tortuous, calcified, or stenosed.
The product stored, handled, inspected and prepped according to the instructions for use (ifu) with no unusual or damages noted about the delivery sheath prior to use.
The device package was not damaged.
Excessive force was not used when the device was removed from the packaging.
There was no excessive force used during the procedure.
The doctor requested a customer letter.
The device is available for analysis.
No other information was provided.
|