(b)(4).Batch # unknown.Device analysis: the analysis results confirmed that the lc310 device was returned nonfunctional as the jaws were misaligned; therefore, the clips could not be loaded into the jaws.The certificate records are accessible through external manufacturing, however, no lot or batch were provided, dhr could not be performed.Please note that as stated in the ifu: "all surgical instruments are subject to a degree of wear and tear because of normal use.A regular and precise visual check of the instrument should be made before each use.".
|